Microscope Heating and Cooling Stage Spectrometer and Integrating Sphere Heating and Cooling Stage Electrical Measurement Heating and Cooling Stage Aerogel Deep Cryogenic Processing Heating and Cooling Stage Mechanical Testing Heating and Cooling Stage Heating and Cooling Stage for Other Applications
-
Applications
-
-
About
-
Microscope Thermal Stage | Spectrometer and Integrating Sphere Thermal Stage | Electrical Test Thermal Stage
Vacuum pA‑level current, four‑probe method, temperature‑controlled stage, on‑site testing
Category: Microelectronics
Release date: 2024-02-13
Summary: Recently, in response to customer requirements, test engineers conducted microscope‑compatible and temperature‑control tests on a picoampere‑level current 4‑probe stage with thermal cycling capability.
Recently, in response to customer requirements, test engineers conducted microscope‑compatibility and temperature‑control tests on a picoampere‑level current four‑probe thermal stage.
I. Product Overview
The Congtical Technology Pi’an‑grade Current‑4‑Probe Temperature‑Controlled Stage is a product specifically designed for variable‑temperature electrical measurements in materials research. It enables characterization of materials during both heating and cooling cycles, allowing measurement of temperature‑dependent resistance and the evolution of electrical properties with temperature.
This model achieves high-precision XYZ‑axis displacement by adjusting the external probe arm to reposition the internal probe. Additionally, it features 24 internal electrode leads, facilitating connection to a variety of source meters for conducting diverse electrical measurements, such as dielectric constant, ferroelectric constant, I‑V characteristics, and temperature‑dependent resistance.
II. Product Specifications
|
Project |
Technical Specifications |
|
Hot and cold methods |
Liquid nitrogen cooling, resistive heating |
|
Temperature range |
-190~500℃ |
|
Temperature stability |
±0.1℃ |
|
Temperature resolution |
0.1℃ |
|
Current accuracy |
1pA |
|
Displacement accuracy |
4-axis DC XYZ positioning accuracy: ±0.01 mm; XYZ travel range: ±6.5 mm. |
|
Vacuum level |
≤0.1 Pa (including mechanical pump and vacuum gauge) |
|
Probe |
4 sets of coaxial probes, vacuum‑compatible triaxial SMA connectors |
|
Stage size |
31×30 mm square stage |
|
Stage material |
Silver-plated |
III. Test Results
RT~500℃
RT~ -190℃
Thermistor Testing