Recently, according to customer needs, test engineers conducted microscope adaptation tests and temperature control tests on the picoamp current 4-probe displacement hot and cold stage.
1. Product introduction
Congtical picoamp current 4-probe displacement hot and cold stage is a product designed for temperature-changing electrical testing in the material research process. It can characterize the heating and cooling stages of materials, test temperature-changing resistance, and the characteristics of electrical properties changing with temperature.
This model can perform high-precision XYZ direction displacement by adjusting the external probe arm and moving the internal probe position. In addition, 24 electrode lines are designed internally to facilitate connection with a variety of source meters for different types of electrical tests (such as dielectric tests). constant, ferroelectric constant, IV curve, variable temperature resistance, etc.).
The vacuum picoamp current 4-probe displacement hot and cold stage uses liquid nitrogen refrigeration and resistance heating to achieve precise control within the range of -190~500°C. It can be integrated with electrical instruments (such as source meters, multimeters, etc.) to perform temperature changes. test.
2. Product parameters
(1)Hot and cold methods: liquid nitrogen refrigeration, resistance heating
(2)Temperature control range: -196~500℃
(3)Temperature stability: ±0.1℃
(4)Temperature resolution: 0.1℃
(5)Current accuracy: 1pA
(6)Vacuum degree ≤ 0.1Pa (including mechanical pump and vacuum gauge)
(7)4-arm DC XYZ control accuracy ±0.01mm, XYZ displacement range ±6.5mm
(8)Probes: 4 sets of coaxial probes, vacuum tri-coaxial SMA interface
(9)Stage: 31*30mm square stage
(10)Stage material: silver
3. Test results
RT~500℃
RT~ -190℃
Variable temperature resistivity test